Setting up the balance
14
B
A
A
2.4 Power supply
Your balance is delivered complete with an AC adapter and a country-specific power supply
cable. The AC adapter is suitable for all power supply voltages in the range:
100 – 240 VAC, -10/+15%, 50/60 Hz.
Check that the local power supply voltage is in this range. If it is not, DO NOT connect
the balance or the AC adapter to the power supply, and contact your METTLER TOLEDO
dealer.
Plug the AC adapter into the socket in the back of your balance (see illustration) and into
the power supply.
Important: Guide the cables so that they cannot become damaged
or interfere with the weighing process! Take care that the AC adapter
cannot come into contact with liquids!
After the balance has been connected to the power supply, it carries out
a self test and is then ready for operation.
Note: If the display field remains dark, even though the power supply
connection functions, first disconnect the balance from the power sup-
ply. Open the terminal by pressing the two buttons (A), then swivel the
upper part of the terminal open and check that the cable (B) is correctly
plugged in.