Technical specifications GPro™ 500 TDL Series
Technical specifications
Measurement
Effective path length 100, 200, and 400 mm (3.94", 7.87", and 15.75")
(EPL) When configuring the GPro 500 with the M400, the double value of the effective
path length must be keyed in (2⫻ effective path length).
Detection limit 0.01% Vol (100 ppm-v) with 1 m path length and ambient standard conditions
(no dust load, dry gas, O
2
in N
2
)
Display units ppm-v, % Vol O
2
Accuracy 2% of reading or 100 ppm, whichever is bigger
Linearity Better than 1%
Resolution <0.01% Vol O
2
(100 ppm-v)
Drift Negligible (<2% of measuring range between maintenance intervals)
Sampling rate 1s
Response time (T
9
0
)O
2
in N
2
21% → 0% in <2s
Repeatability ± 0.25 % of reading or 0.05 % O
2
, whichever is greater
Electrical inputs & outputs
Communication interface RS485 (to transmitter) or direct current outputs (optional)
Service interface Ethernet (to PC) as direct service interface for FW updates (not using the M400
transmitter), for off-line diagnostics and configuration database up/download
Memory slot interface SD card reader/writer for data retrieval (measurement & diagnostics), FW update
and remote diagnostics (configuration file up/download) (to be accessed using
Ethernet port). Space for data storage: 4 GB.
Analog outputs (on M400) 4⫻ 4– 20 mA (22 mA): process temperature, pressure, % conc, % transmission
Number of direct 2 (optional)
analog outputs
Current outputs Passive 4…20 mA outputs, galvanically isolated, alarms to 3.6 mA or 22 mA
conform to NAMUR NE43 guidelines
Measurement error Non-linearity <±0.002 mA over the 1 to20 mA range
through analog outputs Offset error <±0.004 mA (zero scale)
Gain error <±0.04 mA (full scale)
Analog output configuration Linear
Load Max 500 Ohms
Hold mode input Yes, via Ethernet (using the MT-TDL suite)
Analog inputs 2⫻ 4– 20 mA (passive) for pressure and temperature (optional: fixed values)
Display On M400, see M400 technical datasheet
Relays 4 relays (on M400)
Power supply 24VDC, 0.2 A, >5 W
Fuse 1 A slow blow type FC
Calibration
Calibration (factory) Full calibration
Calibration (user) One-point and process calibration